Volume 18-19

Aqueous standard solutions of ethanol for calibration of breath simulators made by gravimetric preparation

Authors Piotr Janko, Robert Kordulasiński, Jolanta Wasilewska, Elżbieta Lenard Central Office of Measures (Główny Urząd Miar)

Abstract

Breath analyzers are instruments for determination of alcohol (ethanol) mass concentration in exhaled air. These instruments are calibrated in Poland by means of wet gas standards produced „in situ” from aqueous standard solutions in breath simulators. The paper presents a procedure developed at GUM for the gravimetric preparation of liquid ethanol standards along with a detailed analysis of the potential sources of uncertainty of assigned ethanol content.

Bibliography

[1] G. Baptista, F. Dias, E. Filipe: Alcohol Breath Analysers: Tracebility Chain in Portugal. From primary gas standards to metrological control. Proceedding of 14eme Congress International de Metrologie, Paris, France, 2009.
[2] DOT NHTSA, Highway Safety Programs; Conforming Products List of Calibrating Units for Breath Alcohol Testers. Federal Register, vol. 77, no. 204, October 22, 2012.
[3] K. M. Dubowski: Breath-Alcohol Simulators, Scientific Basis and Actual Performance. J. Anal. Toxicol., vol. 3 (1979), p. 177-182.
[4] International Rcommendations. Evidential breath analyzers. OIML R126:2012.
[5] Rozporządzenie Ministra Gospodarki z dnia 25 maja 2006 r. w sprawie liczbowych danych odniesienia dla mieszanin alkoholu etylowego i wody. Dz.U. nr 106 poz. 716, 2006. International Alcoholometric Tables, OIML R22:1973.
[6] P. De Bièvre, R. Dybkær, A. Fajgelj, D. B. Hibbert: Metrological traceability of measurement results in chemistry: Concepts and implementation (IUPAC Technical Report). Pure Appl. Chem., vol. 83 (2011).
[7] B. King, R. E. Lawn: International Interlaboratory Study of Forensic Ethanol Standards. CITAC Project Report, 1997.
[8] B. King, R. Lawn: International interlaboratory study of forensic ethanol standards. Analyst, vol. 124 (1999), p. 1123-1130,.
[9] N. Chang et al.: Preparation, Uncertainty, & Certification of Ethanol Standards. Cerillant.
[10] M. Archer, B. J. de Vos, M. S. Visser: The preparation, assay and certification of aqueous ethanol reference solutions. Accreditation Qual. Assur., vol. 12 (2007), p. 188-193.
[11] National Institute of Standards & Technology. Certificate of Analysis. Standard Reference Material 2891. Ethanol-Water Solution (Nominal Mass Fraction 0.02 %). NIST, 2013.
[12] Certificate of Analysis. ERM - AC510a. Aqueous ethanol 50 mg/100ml. LGC, 2015.
[13] R. Philipp, O. Hanebeck, S. Hein, W. Bremser, T. Win, I. Nehls: Ethanol/water solutions as certified reference materials for breath alcohol analyzer calibration. Accreditation Qual. Assur., vol. 15 (2010) p. 141-146.
[14] PN-EN ISO 3696:1999: Woda stosowana w laboratoriach analitycznych – Wymagania i metody badań.
[15] A. Picard, R. S. Davis, M. Gläser, K. Fujii: Revised formula for the density of moist air (CIPM-2007). Metrologia, vol. 45 (2008), p. 149-155.
[16] M. Tanaka, G. Girard, R. Davis, A. Peuto, N. Bignell: Recommended table for the density of water between 0 °C and 40 °C based on recent experimental reports. Metrologia, vol. 38 (2001), p. 301.
[17] P. Janko, R. Kordulasiński: Badanie wpływu higroskopijności etanolu na wartość ułamka masowego etanolu i jego niepewność w sporządzanych grawimetrycznie wodnych roztworach wzorcowych stosowanych w symulatorze wydechu. Sprawozdanie GUM, 2015.
[18] R. N. Kacker, J. F. Lawrence: Trapezoidal and triangular distributions for Type B evaluation of standard uncertainty. Metrologia, vol. 44 (2007), p. 117-127.
[19] D. B. G. Williams, M. Lawton: Drying of Organic Solvents: Quantitative Evaluation of the Efficiency of Several Desiccants. J. Org. Chem., vol. 75 (2010), p. 8351-8354.
[20] K. M. Dubowski, E. E. Goodson, M. Sample: Storage stability of simulator ethanol solutions for vapor-alcohol control tests in breath-alcohol analysis. J. Anal. Toxicol., vol. 26 (2002), p. 406-410.
[21] B. L. C. Chow, J. G. Wigmore: The Stability of Aqueous Alcohol Standard used in Breath Alcohol Testing after Twenty-Six Years Storage. Can. Soc. Forensic Sci. J., vol. 38 (2005), p. 21-24.
[22] V. de Souza, J. M. R. Caixeiro, R. D. C. Cunha, L. A. N. Valente, R. M. H. Borges, and R. A. L. da Silva: Evaluation Of Stability Of Ethanol In Water Certified Reference Material: Measurement Uncertainty Under Transport And Storage Conditions, presented at the XVIII IMEKO WORLD CONGRESS. Metrology for a Sustainable Development, Rio de Janeiro, Brazil, 2006.
[23] V. de Souza, J. M. Rodrigues, R. D. da C. Cunha Bandeira, L. A. das Neves Valente, M. V. Barreto Sousa, V. F. da Silva, R. A. Lemos da Silva: Evaluation of the stability of ethanol in water certified reference material: measurement uncertainty under transport and storage conditions. Accreditation Qual. Assur., vol. 13 (2008), p. 717-721.
[24] J. Kucmanic: Long-Term Stability of Ethanol Solutions for Breath-Alcohol Tests. J. Anal. Toxicol., vol. 33 (2009), p. 328-331.
[25] Evaluation of the Uncertainty of Measurement in calibration. EA-4/02 M: 2013.